Comply with standard IEC 60711,ITU-T P.57
- Product Detail
- Technical specifications
Model | AWA6160 | AWA6161 | AWA6162 | AWA6163 |
Name | Ear Simulator | 2cm3 Coupler | Ear Simulator | Ear Simulator (6 cc) |
Fulfils Standard | IEC 60318-1 (IEC 318-1970) | IEC60318-5 (IEC60126 -1973) | IEC 60711, ITU-T P.57 | IEC60318-3 (IEC60303 -1970) |
Frequency Range | 20Hz~10,000Hz | 125 Hz~8,000 Hz | 20 Hz~10,000 Hz | 100 Hz~8,000 Hz |
Frequency Response | ≤±1 dB | ≤±0.5dB(200Hz~5,000Hz)
≤±1dB(125 Hz~8,000Hz) |
≤ ±0.5dB(200Hz~5,000 Hz)
≤±1 dB(20 Hz~10,000 Hz) |
≤ ±0.5dB(200Hz~5,000 Hz)
≤±1 dB(100 Hz~8,000 Hz) |
Used with Microphone | Model AWA14422 1/2 inch pressure field Microphone | Model AWA14422 1/2 inch pressure field Microphone, also 1-inch pressure field Microphone | Model AWA14424D 1/2 inch pressure field Microphone | Model AWA14412 1-inch pressure field Microphone |
Used with Preamplifier | Model AWA14604 preamplifier or AWA14601 preamplifier | |||
Operation Voltage Range for the Preamplifier | 12 V~36 V | |||
Length of cable | 2 m | |||
Accessories | Ear Simulator fixed block | Coupler fixed block,
Adaptor for 1/2 inch to 1 inch |
Ear Simulator fixed block | Ear Simulator fixed block |
Uses | For calibrating supra-aural earphones used in audiometry and telephone -entry applied to the ear without acoustical leakage. | Frequency response, sensitivity, and distortion measurements on insert earphones, e.g. used in hearing aids, operator headsets, etc. | For measurements of air conduction hearing aids and earphones, are coupled to the ear by means of ear inserts e.g. ear moulds or similar devices. | For calibrating audiometers and audiometry headphone |
Prev: AWA6161 Ear Simulator
Next: AWA6163 Ear Simulator